Exploring the Advantages of Blind Source Separation in Monitoring Input Respiratory Impedance During Apneic Events

Abstract

In this contribution, we investigate the effectiveness of using filtering techniques based on blind source separation methods for continuous monitoring of input respiratory impedance to detect obstructive sleep apnea (OSA). The simple and non-invasive forced oscillation lung function test is used to apply respectively a 1Hz and a 3Hz input pressure oscillation to the patient during spontaneous ventilation. Once the breathing signal extracted by means of independent component analysis (FastICA) method, the input respiratory impedance can be obtained by the ratio between respiratory pressure and flow at the excited frequency. This index is then used to detect changes in the respiratory impedance values as a result of the apneic events. The advantages of the blind source separation techniques are emphasized and its use to assess respiratory mechanics for detecting OSA episodes is extracted from this study.